Calibration of moisture monitors [electronic resource].
- Published:
- Los Alamos, N.M. : Los Alamos Scientific Laboratory, 1979.
Oak Ridge, Tenn. : Distributed by the Office of Scientific and Technical Information, U.S. Dept. of Energy. - Physical Description:
- Pages: 10 : digital, PDF file
- Additional Creators:
- Los Alamos Scientific Laboratory
United States. Department of Energy. Office of Scientific and Technical Information - Access Online:
- www.osti.gov
- Summary:
- A method for calibrating an aluminum oxide hygrometer against an optical chilled mirror dew-point hygrometer has been established. A theoretical cross-point line of dew points from both hygrometers and a maximum moisture content of 10 ppM/sub v/ are used to define an area for calibrating the sensor probes of the aluminum oxide hygrometer.
- Subject(s):
- Note:
- Published through SciTech Connect.
02/01/1979.
"la-7696-ms"
Gutierrez, R.L. - Funding Information:
- W-7405-ENG-36
View MARC record | catkey: 14117527