Large-area PSPMT based gamma-ray imager with edge reclamation [electronic resource].
- Published:
- Washington, D.C. : United States. Dept. of Energy, 2000.
Oak Ridge, Tenn. : Distributed by the Office of Scientific and Technical Information, U.S. Dept. of Energy. - Physical Description:
- PDF-file: 11 pages; size: 7 Mbytes
- Additional Creators:
- Lawrence Berkeley National Laboratory
United States. Department of Energy
United States. Department of Energy. Office of Scientific and Technical Information - Access Online:
- www.osti.gov
- Summary:
- We describe a coded aperture, gamma-ray imager which uses a CsI(Na) scintillator coupled to an Hamamatsu R3292 position-sensitive photomultiplier tube (PSPMT) as the position-sensitive detector. We have modified the normal resistor divider readout of the PSPMT to allow use of nearly the full 10 cm diameter active area of the PSPMT with a single scintillator crystal one centimeter thick. This is a significant performance improvement over that obtained with the standard readout technique where the linearity and position resolution start to degrade at radii as small as 3.5 cm with a crystal 0.75 crn thick. This represents a recovery of over 60% of the PSPMT active area. The performance increase allows the construction of an imager with a field of view 20 resolution elements in diameter with useful quantum efficiency from 60-700 keV. In this paper we describe the readout technique, its implementation in a coded aperture imager and the performance of that imager.
- Note:
- Published through SciTech Connect.
09/21/2000.
"ucrl-jc-138661"
IEEE Transactions on Nuclear Science
2000 Institute of Electrical and Electronics Engineers Nuclear Science Symposium and Medical Imaging Conference, Lyon, France, Oct 15 - Oct 20, 2000.
Nakae, L; Ziock, K-P. - Funding Information:
- W-7405-ENG-48