Ion Beam Induced Charge Collection (IBICC) Studies of Cadmium Zinc Telluride (CZT) Radiation Detectors [electronic resource].
- Published:
- Washington, D.C. : United States. Dept. of Energy, 1999.
Oak Ridge, Tenn. : Distributed by the Office of Scientific and Technical Information, U.S. Dept. of Energy. - Physical Description:
- 16 pages : digital, PDF file
- Additional Creators:
- Sandia National Laboratories, United States. Department of Energy, and United States. Department of Energy. Office of Scientific and Technical Information
Access Online
- Restrictions on Access:
- Free-to-read Unrestricted online access
- Summary:
- Cadmium Zinc Telluride is an emerging material for room temperature radiation detectors. In order to optimize the performance of these detectors, it is important to determine how the electronic properties of CZT are related to the presence of impurities and defects that are introduced during the crystal growth and detector fabrication. At the Sandia microbeam facility IBICC and Time Resolved IBICC (TRIBICC) were used to image electronic properties of various CZT detectors. Two-dimensional areal maps of charge collection efficiency were deduced from the measurements. In order to determine radiation damage to the detectors, we measured the deterioration of the IBICC signal as the function of dose. A model to explain quantitatively the pattern observed in the charge collection efficiency maps of the damaged detectors has been developed and will be discussed in the paper.
- Report Numbers:
- E 1.99:sand99-0917c
sand99-0917c - Subject(s):
- Other Subject(s):
- Note:
- Published through SciTech Connect.
07/22/1999.
"sand99-0917c"
International Conference on Ion Beam Analysis, Dresden (DE), 07/26/1999--07/31/1999.
Doyle, B.L.; Walsh, D.S.; Vizkelethy, G. - Funding Information:
- AC04-94AL85000
View MARC record | catkey: 14400164