Actions for Micro mass spectrometer on a chip [electronic resource].
Micro mass spectrometer on a chip [electronic resource].
- Published
- Washington, D.C. : United States. Dept. of Energy, 2005.
Oak Ridge, Tenn. : Distributed by the Office of Scientific and Technical Information, U.S. Dept. of Energy. - Physical Description
- 182 pages : digital, PDF file
- Additional Creators
- Sandia National Laboratories, United States. Department of Energy, and United States. Department of Energy. Office of Scientific and Technical Information
Access Online
- Restrictions on Access
- Free-to-read Unrestricted online access
- Summary
- The design, simulation, fabrication, packaging, electrical characterization and testing analysis of a microfabricated a cylindrical ion trap (μCIT) array is presented. Several versions of microfabricated cylindrical ion traps were designed and fabricated. The final design of the individual trap array element consisted of two end cap electrodes, one ring electrode, and a detector plate, fabricated in seven tungsten metal layers by molding tungsten around silicon dioxide (SiO₂) features. Each layer of tungsten is then polished back in damascene fashion. The SiO₂ was removed using a standard release processes to realize a free-hung structure. Five different sized traps were fabricated with inner radii of 1, 1.5, 2, 5 and 10 {micro}m and heights ranging from 3-24 {micro}m. Simulations examined the effects of ion and neutral temperature, the pressure and nature of cooling gas, ion mass, trap voltage and frequency, space-charge, fabrication defects, and other parameters on the ability of micrometer-sized traps to store ions. The electrical characteristics of the ion trap arrays were determined. The capacitance was 2-500 pF for the various sized traps and arrays. The resistance was in the order of 1-2 Ω. The inductance of the arrays was calculated to be 10-1500 pH, depending on the trap and array sizes. The ion traps' field emission characteristics were assessed. It was determined that the traps could be operated up to 125 V while maintaining field emission currents below 1 x 10⁻¹⁵ A. The testing focused on using the 5-{micro}m CITs to trap toluene (C₇H₈). Ion ejection from the traps was induced by termination of the RF voltage applied to the ring electrode and current measured on the collector electrode suggested trapping of ions in 1-10% of the traps. Improvements to the to the design of the traps were defined to minimize voltage drop to the substrate, thereby increasing trapping voltage applied to the ring electrode, and to allow for electron injection into, ion ejection from, and optical access to the trapping region.
- Report Numbers
- E 1.99:sand2005-6838
sand2005-6838 - Subject(s)
- Other Subject(s)
- Note
- Published through SciTech Connect.
11/01/2005.
"sand2005-6838"
Blain, Matthew Glenn; Cruz, Dolores Y.; Fleming, James Grant. - Funding Information
- AC04-94AL85000
View MARC record | catkey: 14674228