Resolution limitations and optimization of the ITT F4157 streak tube focus for fast (10 ps) operation [electronic resource].
- Published
- Livermore, Calif : Lawrence Livermore National Laboratory, 1987.
Oak Ridge, Tenn. : Distributed by the Office of Scientific and Technical Information, U.S. Dept. of Energy. - Physical Description
- Pages: 22 : digital, PDF file
- Additional Creators
- Lawrence Livermore National Laboratory and United States. Department of Energy. Office of Scientific and Technical Information
Access Online
- Restrictions on Access
- Free-to-read Unrestricted online access
- Summary
- The ITT F4157 image tube is biased at voltages far from the original design for operation in an ultrafast (10 ps) streak camera. Its output resolution at streak camera operating potentials has been measured as a function of input slit width, incident-light wavelength, and focus-grid voltage. The results are similar to those reported for the RCA C73435 streak tube. Indeed, the two tubes can be substituted for each other with minor mechanical modifications. The temporal resolution is insensitive to focus-grid voltage for a narrow (50 ..mu..m) input slit, but is very sensitive to focus-grid voltage for a wide (500 ..mu..m) input slit. Spatial resolution is nearly independent of focus-grid voltage for values that give good temporal resolution. Both temporal and spatial resolution depend on the incident-light wavelength. Streak camera operation is simulated with a computer program that calculates photoelectron trajectories. Electron ray tracing describes the observed effects of slit width, incident-light wavelength, and focus-grid voltage on the output resolution. 6 refs., 10 figs., 1 tab.
- Report Numbers
- E 1.99:ucid-21258