CMM Interim Check Design of Experiments (U) [electronic resource].
- Published
- Washington, D.C. : United States. Dept. of Energy, 2015.
Oak Ridge, Tenn. : Distributed by the Office of Scientific and Technical Information, U.S. Dept. of Energy - Physical Description
- 25 pages : digital, PDF file
- Additional Creators
- Los Alamos National Laboratory, United States. Department of Energy, and United States. Department of Energy. Office of Scientific and Technical Information
Access Online
- Restrictions on Access
- Free-to-read Unrestricted online access
- Summary
- Coordinate Measuring Machines (CMM) are widely used in industry, throughout the Nuclear Weapons Complex and at Los Alamos National Laboratory (LANL) to verify part conformance to design definition. Calibration cycles for CMMs at LANL are predominantly one year in length and include a weekly interim check to reduce risk. The CMM interim check makes use of Renishaw’s Machine Checking Gauge which is an off-the-shelf product simulates a large sphere within a CMM’s measurement volume and allows for error estimation. As verification on the interim check process a design of experiments investigation was proposed to test a couple of key factors (location and inspector). The results from the two-factor factorial experiment proved that location influenced results more than the inspector or interaction.
- Report Numbers
- E 1.99:la-ur--15-25946
la-ur--15-25946 - Subject(s)
- Note
- Published through SciTech Connect.
07/29/2015.
"la-ur--15-25946"
Montano, Joshua Daniel [Los Alamos National Laboratory]. - Funding Information
- AC52-06NA25396
View MARC record | catkey: 23766851