Actions for Squid based beam current meter
Squid based beam current meter
- Author
- Kuchnir, M.
- Published
- United States : [publisher not identified], 1983
Springfield, Va.: National Technical Information Service, [approximately 1983] - Physical Description
- microfiche : negative ; 11 x 15 cm
- Summary
- A SQUID based beam current meter has the capability of measuring the current of a beam with as little as 30 x 155 antiprotons (with a signal to noise ratio of 2). If low noise dc current is used to cancel most of the beam or an up-down counter is used to count auto-resets this sensitivity will be available at any time in the acumulation process. This current meter will therefore be a unique diagnostic tool for optimizing the performance of several Tev I components. Besides requiring liquid helium it seems that its only drawback is not to follow with the above sensitivity a sudden beam change larger than 16 ..mu..A, something that could be done using a second one in a less sensitive configuration.
- Report Numbers
- DE84004777; FERMILAB/TM-1225
- Other Subject(s)
- 43 particle accelerators
- 430303 - particle accelerators- experimental facilities & equipment
- Antinucleon beams
- Antiparticle beams
- Antiproton beams
- Beam currents
- Beam monitors
- Beams
- Cost
- Currents
- Design
- Electronic equipment
- Equipment
- Fluxmeters
- Measuring instruments
- Microwave equipment
- Monitors
- Specifications
- Squid devices
- Superconducting devices
- Collection
- NTIS collection.
- Note
- DOE contract number: AC02-76CH03000
OSTI Identifier 5481159
Research organization: Fermi National Accelerator Lab. (FNAL), Batavia, IL (United States).
View MARC record | catkey: 47336401