Actions for Impact of instrument response variations on health physics measurements
Impact of instrument response variations on health physics measurements
- Author
- Armantrout, G. A.
- Published
- United States : [publisher not identified], 1984
Springfield, Va.: National Technical Information Service, [approximately 1984] - Physical Description
- microfiche : negative ; 11 x 15 cm
- Summary
- Uncertainties in estimating the potential health impact of a given radiation exposure include instrument measurement error in determining exposure and difficulty in relating this exposure to an effective dose value. Instrument error can be due to design or manufacturing deficiencies, limitations of the sensing element used, and calibration and maintenance of the instrument. This paper evaluates the errors which can be introduced by design deficiencies and limitations of the sensing element for a wide variety of commonly used survey instruments. The results indicate little difference among sensing element choice for general survey work, with variations among specific instrument designs being the major factor. Ion chamber instruments tend to be the best for all around use, while scintillator-based units should not be used where accurate measurements are required. The need to properly calibrate and maintain an instrument appears to be the most important factor in instrument accuracy. 8 references, 6 tables.
- Report Numbers
- DE85000860; UCRL-91682; CONF-841007-7
- Other Subject(s)
- Collection
- NTIS collection.
- Note
- DOE contract number: W-7405-ENG-48
OSTI Identifier 6315141
Research organization: Lawrence Livermore National Lab., CA (USA).
View MARC record | catkey: 47345893